High Performance Oscilloscope Calibrator
Routine Maintenance
3
Once the cover is removed, holes can be seen in the chassis assembly right side, with the
reticulated filter located over these holes, attached by four black nylon snap rivets
through four holes to the chassis assembly.
Inspect and clean the filter element by vacuuming at intervals of no more than 90 days.
Remove the filter for thorough cleaning at least once per year, immediately prior to
routine calibration of the unit.
Removing the Top Cover
W
Warning
To prevent fire, explosion, or personal injury, disconnect the
power line cord before removing any covers.
The side edges of the top cover are located in slots in both side extrusions, and secured to
the rear panel by two screws. A solid plastic block protects each of the rear corners.
To access the filter element, remove the top cover as follows:
1.
Release the four screws securing the two rear-corner blocks and remove the
blocks.
2.
Release the two screws securing the top cover to the rear panel.
3.
Pull the top cover to the rear to clear the front bezel, then lift off to the rear.
Top Guard Shield and Calibration Seal
W
Caution
With the cover removed, a 'Calibration seal' can be seen
covering a countersunk screwhead on the front of the top guard
shield. The seal is set in position following a calibration of the
instrument, so that removal of the guard shield can be detected.
The calibration seal must NOT be broken unless the guard
shield is to be removed for Authorized work inside the chassis
assembly.
Removal of the guard shield will compromise the traceable
calibration of the instrument, and a full recalibration of the
9500B will be required. If removal of the guard shield is
authorized, take care to heed the WARNING printed on it next to
the
XW
symbols.
Removing the Filter Element
The filter is removed as follows:
Carefully lever out the center pin of each of the four black nylon snap rivets and pull out
the snap latches. Lift off the filter element.
Cleaning the Filter Element
Once the filter is removed, wash it in warm water and household detergent, rinse
thoroughly and allow to dry.