9500B
Calibration Manual
138
12.
The measured edge rise time value is the average of 8 separate measurements.
Use the following sequence for measuring the remaining measurement values.
a)
Press the calibrator OUTPUT OFF button.
b)
Press the calibrator OUTPUT ON button.
c)
Repeat the measurement and record the measured rise time.
d)
Repeat this sequence of steps until you have recorded a total of 8
measurements.
13.
Calculate the average of these 8 measurements to a precision of 2 decimal places.
This is the calculated average rise time of the combined 9550 signal rise time and
the oscilloscope response rise time.
14.
Using the following formula, determine the 9550’s calibrated edge rise time by
subtracting the measured (or calibrated) oscilloscope measurement response rise
time from your calculated average.
Edge Rise Time =
15.
Save this calculated edge rise time value for later use in calibrating the head.
16.
Prepare to measure the 9550’s falling edge transition time by selecting the falling
edge with the soft key. On the 9500B confirm operation with the settings at an
Amplitude of 500 mV, Frequency at 1MHz, Falling Edge, and OUTPUT ON.
17.
Repeat steps 8 to 15 for the 9550’s falling edge time measurement.
18.
Turn the OUTPUT OFF.
Save Calibration Data to Active Head
1.
Insure the rear-panel Calibration Enable switch is in the “ENABLE” position.
2.
Press the MODE function key to access the main menu screen.
3.
Press the CALB Key on the main menu.
4.
Enable changing calibration constants by entering a valid password to access the
Calibration Mode display screen.
5.
From the main menu screen selection, select the HEAD CAL Mode of operation.
6.
Confirm the CHANNEL SELECT settings are appropriate, then select EDGE
and 25 ps.
7.
Select TARGET 1.
8.
Refer to the Rising Edge speed as calculated in the previous section and enter this
value into the edge speed field on the display screen. Press ACCEPT CAL.