9500B
Calibration Manual
134
Table 41. Edge Function: 150 ps Edge: Speed
Cal. Point
Voltage
Frequency
Edge
Target 1
3.0000 V
1 MHz
Rising
Target 2
2.0000 V
1 MHz
Rising
Target 3
1.0000 V
1 MHz
Rising
Target 4
1.0072 V
1 MHz
Rising
Target 5
316.49 mV
1 MHz
Rising
Target 6
316.49 mV
1 MHz
Rising
Target 7
3.0000 V
1 MHz
Falling
Target 8
2.0000 V
1 MHz
Falling
Target 9
1.0000 V
1 MHz
Falling
Target 10
1.0072 V
1 MHz
Falling
Target 11
316.49 mV
1 MHz
Falling
Target 12
316.49 mV
1 MHz
Falling
Calibration Procedure: 70 ps Edge: Linearity
1.
Ensure that the 9500B is connected to the oscilloscope as shown in Figure 20.
and that both instruments are powered on and warmed up.
2.
Select the required 9500B Trigger Channel, Cable Select, and Trigger Ratio
settings.
3.
Select the required measurement device function to measure edge response.
4.
Ensure that the 9500B is in HEAD CAL, Edge, 70 ps Edge, LIN mode.
5.
Set 9500B's output ON, selecting TARGET 1 from Table 42.
6.
Select a measurement device range that gives an on-scale reading.
7.
Adjust the 9500B's output amplitude to give a reading equal to Target Amplitude
on the oscilloscope.
8.
Press ACCEPT CALIB.
9.
Select the next TARGET in Table 42 by pressing NEXT TARGET and return to
step 4; repeat until no TARGETs remain.