High Performance Oscilloscope Calibrator
9510/9530/9550/9560 Head Verification by Functions
55
Verifying the Leveled Sine Function: LF Gain
Summary
Equipment requirements are given in Equipment Requirements and test interconnections
in Interconnections.
The Leveled Sine Function is verified by carrying out measurements of amplitude at
frequencies of 1 kHz, 45 kHz and 50 kHz; in the sequences given in Verification Setup
and Verification Procedure, at the verification points shown in Table 11.
Equipment Requirements
•
The UUT Active Head, connected to a verified Model 9500B Mainframe (refer
to Verifying the LF Sine Voltage Function).
•
A high resolution Standards DMM with RMS ACVoltage accuracy of 0.3 % or
better, from 10 mV to 1.5 V, at 1 kHz, 45 kHz and 50 kHz.
Example: Model 1281 Digital Multimeter.
•
An adaptor to convert from PC3.5 or BNC to 4mm leads. Example: Model 4955
Calibration Adaptor and PC3.5 to BNC adaptor.
•
Short, high-quality 4 mm leads for connection between the Calibration Adaptor
the DMM.
Interconnections
erw207
Figure 9. LF Sine Voltage Verification — Interconnections