9500B
Calibration Manual
12
Pressing the NEXT FAILURE screen key will have no further effect.
Printing the Test Results
The PRINT screen key is present on the 'ABORTED' or 'completed' screens after the test
has run to completion or has been aborted. Pressing the PRINT key will print out all the
available results.
Printing will only be possible if a suitable printer is set up, connected and on line. Refer
to Printing Selftest Results.
Selftest at Power-On
Note
Certain catastrophic 'System Trip Errors' may cause the display to flash on
and off, at the point of setting power on. In this case, immediately switch
Power Off and report the fault to your Service Center.
The first normal action at power on is to show the Fluke Calibration logo and Product
name, and then the 9500B will run a fast selftes.
If System Trip failures are encountered the 9500B will show the System Trip screen
('RESUME').
If non-trip failures are encountered the 9500B will lapse into Test mode at the following
stage:
erw177
Subsequent action to view the failures and print the results follows as for a Fast selftest
which has run to completion (Selftest Runs to Completion and Viewing the Test Results).
EXIT will return to the 'Select test' menu screen, where, to return to the 'Mode Selection'
menu screen, press the front panel Mode key.
PRINTcan be used to print out the results of the power-on Fast test.