Calibration Mode Screen Softkeys
The following screen softkeys are active:
SPECIAL
Accesses the Special Calibration menu so that you can perform several 'Chse'
(Characterize) calibration operations, which must be done immediately before carrying
out routine recalibration of the Model 9500B.
FACTORY USE ONLY
Accesses 'factory-set' calibration operations which can only be entered by a special
password. These calibration operations only need to be performed when the instrument is
manufactured, or after certain types of repair have been carried out on it. The password is
not given in this manual; any user who suspects that such calibration may be needed
should contact their Fluke Calibration Service Center.
STD CAL
Accesses the user control screens for routine external Mainframe calibration of each
Model 9500B function able to be calibrated.
Special Calibration
Selecting Special Calibration
Pressing the SPECIAL softkey in the Calibration Mode screen transfers to the Special
Calibration screen: