Electrical Safety Analyzer
Detailed Specifications
71
Class A: Equipment is suitable for use in all establishments other than domestic and those directly connected to a low voltage
power supply network which supplies buildings used for domestic purposes. There may be potential difficulties in ensuring
electromagnetic compatibility in other environments, due to conducted and radiated disturbances.
Caution: This equipment is not intended for use in residential environments and may not provide adequate protection to radio
reception in such environments
Korea (KCC) ........................................................ Class A Equipment (Industrial Broadcasting & Communication Equipment)
Class A: Equipment meets requirements for industrial electromagnetic wave equipment and the seller or user should take notice
of it. This equipment is intended for use in business environments and not to be used in homes.
Detailed Specifications
Test Standard Selections ........................................ ANSI/AAMI ES-1, IEC62353, IEC60601-1, and AN/NZS 3551
Voltage
Ranges (Mains voltage) ...................................... 90.0 to 132.0 V ac rms
180.0 to 264.0 V ac rms
Range (Point-to-point voltage) ............................ 0.0 to 300.0 V ac rms
Accuracy..............................................................
±
(2 % of r 0.2 V)
Earth Resistance
Modes.................................................................. Two terminal
Test Current ........................................................ >200 mA ac
Range .................................................................. 0.000 to 2.000
Ω
Accuracy..............................................................
±
(2 % of r 0.015
Ω
)
Equipment Current
Range .................................................................. 0.0 to 20.0 A ac rms
Accuracy..............................................................
±
(5 % of r (2 counts or 0.2 A, whichever is greater))
Duty cycle ............................................................ 15 A to 20 A, 5 min. on/5 min. off
10 A to 15 A, 7 min. on/3 min. off
0 A to 10 A, continuous
Summary of Contents for ESA612
Page 6: ......
Page 10: ...ESA612 Users Manual iv ...
Page 12: ...ESA612 Users Manual vi ...
Page 26: ...ESA612 Users Manual 12 fis113 png Figure 5 DUT Connections to the Analyzer ...
Page 40: ...ESA612 Users Manual 26 faw19 png Figure 15 Mains to Applied Parts Insulation Test Schematic ...
Page 66: ...ESA612 Users Manual 52 gtv32 png Figure 29 Differential Leakage Current Test Schematic ...
Page 68: ...ESA612 Users Manual 54 fis120 png Figure 30 1 to 10 Adapter Connections ...
Page 70: ...ESA612 Users Manual 56 fis121 png Figure 31 ECG Lead Connection with 1 to 10 Adapter ...
Page 74: ...ESA612 Users Manual 60 fis115 png Figure 34 ECG Monitor Connections ...
Page 88: ......