Electrical Safety Analyzer
Making Point-To-Point Measurements
57
Making Point-To-Point Measurements
The Analyzer can make voltage, resistance, and low
current measurements through its Point-to-Point function.
To access the Point-to-Point function menu shown in
Figure 32, press
. Softkeys F1 through F3 are used
to select the measurement function.
fis108.jpg
Figure 32. Point-To-Point Function Menu
Measuring Voltage
To make a voltage measurement:
1. Press the softkey labeled
Voltage
from the
Point-To-Point menu.
2. Insert test leads in the RED (V/
Ω
/A ) and BLACK
jacks.
3. Place the probe tips across the unknown voltage and
read the measurement in the Analyzer’s display.
The Analyzer will measure up to 300 volts ac.
Measuring Resistance
To make a resistance measurement:
1. Press the softkey labeled
Resistance
from the
Point-To-Point menu.
2. Insert test leads in the RED (V/
Ω
/A ) and BLACK
jacks.
3. Null lead resistance by shorting the leads together
and pressing the softkey labeled
Zero Leads
.
4. Place the probes across the unknown resistance and
read the measurement in the Analyzer’s display.
The Analyzer will measure resistances up to 2.0
Ω
.
Summary of Contents for ESA612
Page 6: ......
Page 10: ...ESA612 Users Manual iv ...
Page 12: ...ESA612 Users Manual vi ...
Page 26: ...ESA612 Users Manual 12 fis113 png Figure 5 DUT Connections to the Analyzer ...
Page 40: ...ESA612 Users Manual 26 faw19 png Figure 15 Mains to Applied Parts Insulation Test Schematic ...
Page 66: ...ESA612 Users Manual 52 gtv32 png Figure 29 Differential Leakage Current Test Schematic ...
Page 68: ...ESA612 Users Manual 54 fis120 png Figure 30 1 to 10 Adapter Connections ...
Page 70: ...ESA612 Users Manual 56 fis121 png Figure 31 ECG Lead Connection with 1 to 10 Adapter ...
Page 74: ...ESA612 Users Manual 60 fis115 png Figure 34 ECG Monitor Connections ...
Page 88: ......