Electrical Safety Analyzer
Simulating ECG Waveforms
59
To select one of the predefined waveforms, press the
softkey labeled
Wave Form.
A scroll box with
next to it
appears above the softkey label. Use
or
to scroll
through the different waveforms.
For all waveforms except VFIB and Triangle, the rate or
frequency of the waveform is adjusted through the softkey
labeled
Frequency
or
Rate
. For some waveforms, there
are more than two frequency or rate selections. For those
waveforms, pressing the softkey labeled
Frequency
or
Rate
will open a scroll box above the softkey label with
next to it. Use
or
to select the frequency or rate. For
those waveforms that have only two frequency or rate
selections, the softkey labeled
Frequency
or
Rate
acts as
a toggle, where each press of the softkey switches to the
other value.