Electrical Safety Analyzer
Performing Electrical Safety Tests
51
Performing a Differential Leakage Current Test
Note
The Differential Leakage Current test is available
when the EN62353 standard is selected.
The differential leakage current test measures the
magnitudes of the differential current flowing in the
Equipment Outlet live and neutral, with power applied to
the equipment outlet. All applied parts should be
connected during this test, if equipment has applicable
applied parts.
To perform a differential leakage current test:
1. Press
.
2. Press the soft key labeled
Differential
.
Figure 29 shows the electrical connections between the
Analyzer and the DUT during a Differential Leakage
Current test.
The following outlet conditions apply when performing this
test:
•
Normal Polarity, Closed Earth
•
Normal Polarity, Open Earth
•
Reversed Polarity, Closed Earth
•
Reversed Polarity, Open Earth
Note
If there are more than five applied parts to
connect to the Analyzer, see Using the 1-to-10
Adapter later in this manual.
Summary of Contents for ESA612
Page 6: ......
Page 10: ...ESA612 Users Manual iv ...
Page 12: ...ESA612 Users Manual vi ...
Page 26: ...ESA612 Users Manual 12 fis113 png Figure 5 DUT Connections to the Analyzer ...
Page 40: ...ESA612 Users Manual 26 faw19 png Figure 15 Mains to Applied Parts Insulation Test Schematic ...
Page 66: ...ESA612 Users Manual 52 gtv32 png Figure 29 Differential Leakage Current Test Schematic ...
Page 68: ...ESA612 Users Manual 54 fis120 png Figure 30 1 to 10 Adapter Connections ...
Page 70: ...ESA612 Users Manual 56 fis121 png Figure 31 ECG Lead Connection with 1 to 10 Adapter ...
Page 74: ...ESA612 Users Manual 60 fis115 png Figure 34 ECG Monitor Connections ...
Page 88: ......