Electrical Safety Analyzer
Performing Electrical Safety Tests
33
Performing a Chassis (Enclosure) Leakage Test
Note
The Chassis (Enclosure) Leakage test is only
available for the IEC60601 or ANSI/AAMI ES1
1993 standard selections.
The Chassis (Enclosure) Leakage Test measures the
current flowing between the DUT’s enclosure and
protective earth. Figure 20 shows the electrical
connections between the Analyzer and the DUT.
To perform a Chassis (Enclosure) Leakage Test:
1. Connect a lead between the Analyzer’s V/
Ω
/A jack
and the DUT’s enclosure.
2. Press the softkey labeled
Chassis
from the Leakage
Current Test menu.
3. The Analyzer displays the measured current.
The Chassis Leakage test can be performed with a
number of fault conditions on the test receptacle. Press
to switch the test receptacle between Normal, Off,
Reverse, and Off. Press
to open and close the
neutral connection to the receptacle. Press
to open
and close the receptacle’s earth connection.
The following outlet conditions apply when performing this
test:
•
Normal Polarity
•
Normal Polarity, Open Earth
•
Normal Polarity, Open Neutral
•
Reversed Polarity
•
Reversed Polarity, Open Earth
•
Reversed Polarity, Open Neutral
IEC60601-1 specifies that the applied parts should be
connected for this measurement. Enable this
measurement by pressing
or
which grounds and
ungrounds all applied parts connection posts.
Summary of Contents for ESA612
Page 6: ......
Page 10: ...ESA612 Users Manual iv ...
Page 12: ...ESA612 Users Manual vi ...
Page 26: ...ESA612 Users Manual 12 fis113 png Figure 5 DUT Connections to the Analyzer ...
Page 40: ...ESA612 Users Manual 26 faw19 png Figure 15 Mains to Applied Parts Insulation Test Schematic ...
Page 66: ...ESA612 Users Manual 52 gtv32 png Figure 29 Differential Leakage Current Test Schematic ...
Page 68: ...ESA612 Users Manual 54 fis120 png Figure 30 1 to 10 Adapter Connections ...
Page 70: ...ESA612 Users Manual 56 fis121 png Figure 31 ECG Lead Connection with 1 to 10 Adapter ...
Page 74: ...ESA612 Users Manual 60 fis115 png Figure 34 ECG Monitor Connections ...
Page 88: ......