Performing a Current Consumption Test
To measure the current consumed by the DUT, press
. The Analyzer displays the current flowing through
the mains connections of the test receptacle.
Performing Leakage Current Tests
The Analyzer measures leakage current for a number of
different DUT configurations. In addition to the leakage
found on the enclosure and the earth connection, the
Analyzer can measure leakage on each connected applied
part and combinations of connected applied parts.
Which leakage tests are available depends on which
standard is selected. See the Selecting the Test Standard
section earlier in this manual to change the standard the
Analyzer is using.
Table 5 lists six leakage current tests that have different
names based on which standard is selected.
Press
to access the leakage current main menu
shown in Figure 18.
Table 5. Test Names Based on Selected Standard
IEC60601
AAMI/NFPA 99
Protective Earth Resistance
Ground Wire Resistance
Earth Leakage Current
Ground Wire Leakage Current
Touch or Enclosure Leakage Current
Chassis Leakage Current
Patient Leakage Current
Lead to Ground Leakage Current
Patient Auxiliary Leakage Current
Lead to Lead Leakage Current
Mains on Applied Part (MAP) Leakage Current
Isolation Leakage Current
Summary of Contents for ESA612
Page 6: ......
Page 10: ...ESA612 Users Manual iv ...
Page 12: ...ESA612 Users Manual vi ...
Page 26: ...ESA612 Users Manual 12 fis113 png Figure 5 DUT Connections to the Analyzer ...
Page 40: ...ESA612 Users Manual 26 faw19 png Figure 15 Mains to Applied Parts Insulation Test Schematic ...
Page 66: ...ESA612 Users Manual 52 gtv32 png Figure 29 Differential Leakage Current Test Schematic ...
Page 68: ...ESA612 Users Manual 54 fis120 png Figure 30 1 to 10 Adapter Connections ...
Page 70: ...ESA612 Users Manual 56 fis121 png Figure 31 ECG Lead Connection with 1 to 10 Adapter ...
Page 74: ...ESA612 Users Manual 60 fis115 png Figure 34 ECG Monitor Connections ...
Page 88: ......