ESA612
Users Manual
16
Setting up the Beeper
To enable or disable the beeper:
1. From the setup menu, press the softkey labeled
More
twice to set F2 to the beeper on/off function.
2. Press the softkey labeled
Beeper
to switch the
beeper on and off.
3. Press the softkey labeled
Done
to go back to the
setup menu.
Viewing Instrument Information
To view information about the Analyzer:
1. From the setup menu, press the softkey labeled
More
twice to set F3 to the instrument information
function.
2. Press the softkey labeled
Instrument Information
.
3. After viewing the displayed information, press the
softkey labeled
Done
to exit the information screen.
Viewing Memory
Refer to the Using Memory section later in the manual to
learn about viewing memory and how to store data in the
Analyzer.
Setting the GFCI Limit
To set the GFCI current limit:
1. From the setup menu, press the softkey labeled
More
to reveal additional menu selections.
2. Press the softkey labeled
Instrument
to reveal the
instrument setup selections.
3. Press the softkey labeled
GFCI Limit
to open the
scroll box above the softkey label.
4. Press
or
to adjust the current limit from 5 to
25 mA.
5. Press the softkey labeled
Done
to exit the GFCI Limit
setup function.
Summary of Contents for ESA612
Page 6: ......
Page 10: ...ESA612 Users Manual iv ...
Page 12: ...ESA612 Users Manual vi ...
Page 26: ...ESA612 Users Manual 12 fis113 png Figure 5 DUT Connections to the Analyzer ...
Page 40: ...ESA612 Users Manual 26 faw19 png Figure 15 Mains to Applied Parts Insulation Test Schematic ...
Page 66: ...ESA612 Users Manual 52 gtv32 png Figure 29 Differential Leakage Current Test Schematic ...
Page 68: ...ESA612 Users Manual 54 fis120 png Figure 30 1 to 10 Adapter Connections ...
Page 70: ...ESA612 Users Manual 56 fis121 png Figure 31 ECG Lead Connection with 1 to 10 Adapter ...
Page 74: ...ESA612 Users Manual 60 fis115 png Figure 34 ECG Monitor Connections ...
Page 88: ......