Electrical Safety Analyzer
Performing Electrical Safety Tests
35
Performing a Lead-to-Ground (Patient) Leakage
Test
Note
The Lead-to-Ground (Patient) Leakage Current
Test is not available for IEC 62353 standard
selections.
The Lead-to-Ground (Patient) Leakage Current test
measures the current flowing between a selected applied
part, selected group of applied parts, or ALL applied parts,
and the Mains PE. Figure 21 shows the electrical
connections between the Analyzer and the DUT.
To perform a Lead-to Ground (Patient) leakage test:
1. Press
.
2. Press the soft key labeled
More
.
3. Select one of the applied part groupings by pressing
or
.
Note
Refer to the testing standard when deciding the
type of the applied parts and how they should be
grouped for testing.
4. Press the soft key labeled
Select
.
5. Press
or
to advance through each applied part
grouping, or the individual applied parts, to ground.
These are selected and measured.
The Lead-to-Ground Leakage test can be performed with
a number of fault conditions on the test receptacle. Press
to switch the test receptacle between Normal, Off,
Reverse, and Off. Press
to open and close the
neutral connection to the receptacle. Press
to open
and close the receptacle’s earth connection.
The following outlet conditions apply when performing this
test:
•
Normal Polarity
•
Normal Polarity, Open Neutral
•
Normal Polarity, Open Earth
•
Reversed Polarity
•
Reversed Polarity, Open Neutral
•
Reversed Polarity, Open Earth
Note
If there are more than five applied parts to
connect to the Analyzer, see Using the 1-to-10
Adapter later in this manual.
Summary of Contents for ESA612
Page 6: ......
Page 10: ...ESA612 Users Manual iv ...
Page 12: ...ESA612 Users Manual vi ...
Page 26: ...ESA612 Users Manual 12 fis113 png Figure 5 DUT Connections to the Analyzer ...
Page 40: ...ESA612 Users Manual 26 faw19 png Figure 15 Mains to Applied Parts Insulation Test Schematic ...
Page 66: ...ESA612 Users Manual 52 gtv32 png Figure 29 Differential Leakage Current Test Schematic ...
Page 68: ...ESA612 Users Manual 54 fis120 png Figure 30 1 to 10 Adapter Connections ...
Page 70: ...ESA612 Users Manual 56 fis121 png Figure 31 ECG Lead Connection with 1 to 10 Adapter ...
Page 74: ...ESA612 Users Manual 60 fis115 png Figure 34 ECG Monitor Connections ...
Page 88: ......