ESA612
Users Manual
18
Performing a Ground Wire (Protective Earth)
Resistance Test
The Ground Wire (Protective-Earth) Resistance test
measures the impedance between the Analyzer’s test
receptacle’s PE terminal and the exposed conductive
parts of the DUT that are connected to the DUT’s
Protective Earth.
Prior to conducting any leakage tests with the Analyzer, it
is best to test the integrity of the ground connection
between the Analyzer’s test receptacle ground and the
DUT’s Protective earth ground or enclosure with this test.
To access the Ground Wire (Protective Earth)
∅
/Null
Resistance Test menu press
.
Notes:
(1) The DUT is powered off for this test.
To perform a ground wire resistance test:
1. Ensure the power cord from the DUT is plugged into
the Analyzer’s test receptacle.
2. Press
to reveal the resistance function menu.
3. Connect one end of a test lead to the V/
Ω
/A jack as
shown in Figure 9.
4. If using an accessories probe, connect it to the other
end of the test lead and place the probe tip into the
∅
/Null
jack. If using an alligator clip accessory,
connect it to the other end of the test lead, place the
null post adapter in the
∅
/Null
jack, and clamp the
alligator clip to the null post adapter.
5. Connect the other end of the test lead to
∅
/Null
jack.
6. Press the softkey labeled
Zero Leads
. The Analyzer
zeroes out the measurement to cancel the test lead
resistance.
7. Connect the test lead coming from the
∅
/Null
jack to
the DUT enclosure or protective earth connection.
8. The measured resistance is displayed as shown in
Figure 9 after the DUT connection(s) is/are made.
fis105.jpg
Figure 9. DUT Ground Resistance Measurement
Summary of Contents for ESA612
Page 6: ......
Page 10: ...ESA612 Users Manual iv ...
Page 12: ...ESA612 Users Manual vi ...
Page 26: ...ESA612 Users Manual 12 fis113 png Figure 5 DUT Connections to the Analyzer ...
Page 40: ...ESA612 Users Manual 26 faw19 png Figure 15 Mains to Applied Parts Insulation Test Schematic ...
Page 66: ...ESA612 Users Manual 52 gtv32 png Figure 29 Differential Leakage Current Test Schematic ...
Page 68: ...ESA612 Users Manual 54 fis120 png Figure 30 1 to 10 Adapter Connections ...
Page 70: ...ESA612 Users Manual 56 fis121 png Figure 31 ECG Lead Connection with 1 to 10 Adapter ...
Page 74: ...ESA612 Users Manual 60 fis115 png Figure 34 ECG Monitor Connections ...
Page 88: ......