ESA612
Users Manual
30
fis117.jpg
Figure 18. Leakage Current Main Menu
Note
The display shown in Figure 18 is the main
leakage current menu when AAMI is the selected
standard.
All leakage currents, with the exception of Lead Isolation
(Mains on Applied parts), are displayed in one of three
ways: AC+DC, AC Only, or DC only. The initial result is
displayed in the appropriate parameter based on the
standard selected. To change the displayed parameter,
press
or
. The present measurement method is
displayed to the right of the current measurement while
leakage current tests are conducted.
Measuring Earth Leakage Current
Note
The Ground Wire (Earth) Leakage test is
available for AAMI, 60601, and not IEC 62353.
To measure the current flowing in the DUT’s protective
earth circuit, press the softkey labeled
Ground Wire
(pending the standard) from the leakage current main
menu. Figure 19 shows the electrical connections between
the Analyzer and the DUT during a Ground Wire Leakage
Current Test.
Within the Ground Wire Leakage Current test there are a
few combination measurements that can be performed.
Pressing
switches the polarity of the mains voltage
applied to the Analyzer’s test receptacle between Normal,
Off, Reverse, and Off. Pressing
opens and closes
the neutral connection to the Analyzer’s test receptacle.
There is no need to open up the test receptacle earth
(ground), since this is done internally during the
measurement.
Summary of Contents for ESA612
Page 6: ......
Page 10: ...ESA612 Users Manual iv ...
Page 12: ...ESA612 Users Manual vi ...
Page 26: ...ESA612 Users Manual 12 fis113 png Figure 5 DUT Connections to the Analyzer ...
Page 40: ...ESA612 Users Manual 26 faw19 png Figure 15 Mains to Applied Parts Insulation Test Schematic ...
Page 66: ...ESA612 Users Manual 52 gtv32 png Figure 29 Differential Leakage Current Test Schematic ...
Page 68: ...ESA612 Users Manual 54 fis120 png Figure 30 1 to 10 Adapter Connections ...
Page 70: ...ESA612 Users Manual 56 fis121 png Figure 31 ECG Lead Connection with 1 to 10 Adapter ...
Page 74: ...ESA612 Users Manual 60 fis115 png Figure 34 ECG Monitor Connections ...
Page 88: ......