Electrical Safety Analyzer
Performing Electrical Safety Tests
43
Performing an Alternative Applied Part Leakage
Test
Note
The Alternative applied part leakage test is
available when the EN62353 standard is
selected.
During the Alternative Applied Part Leakage test, the test
voltage is applied between short-circuited applied parts of
a single function and the short-circuited equipment outlet
mains live, neutral, equipment outlet earth, and exposed
conductive surface on the housing. This test should only
be done for equipment with F-Type applied parts. For
equipment with multiple applied parts, test each group of
applied parts of a single function in turn with all others
floating during the test
.
All applied parts can be connected
to the Analyzer’s applied parts jacks and the lead selection
will float those not selected.