6 Mounting
6.2 Transducers
FLUXUS H721
2022-05-15, UMFLUXUS_H721V1-5EN
62
• Hold the tabs of the clip using pliers and pull the solid band completely through the slot of the clip, see Fig. 6.73.
• Bend the end of the solid band back over the set of teeth, see Fig. 6.74.
• Fold the tabs of the clip over the solid band and cut the protruding band off, see Fig. 6.75.
• Position the tension strap clamp on the opposite side of the clip.
• Measure the distance x, see Fig. 6.76. Repeat the steps for the second band and place it at the measured distance x to
the other.
• Measure the distance x, see Fig. 6.76. Repeat the steps for the second tension strap and attach it to the tube at ate
measured distance x.
• Mount the PermaLok onto the tension strap clamp and tighten the screws.
• Insert the transducers into the PermaLok as described in the section 6.2.2.6.
Fig. 6.73
Fig. 6.74
Fig. 6.75
Fig. 6.76
1 – tension strap clamp
2 – rail
3 – solid band
2
3
1
x