![Flexim FLUXUS H721 Operating Instruction Download Page 190](http://html1.mh-extra.com/html/flexim/fluxus-h721/fluxus-h721_operating-instruction_2304735190.webp)
19 Advanced functions
19.8 Event protocol
FLUXUS H721
2022-05-15, UMFLUXUS_H721V1-5EN
190
19.8
Event protocol
If an error occurs, an error message will be displayed in the first line indicating the symbol
. The error message can be
displayed via the menu item
Event log
.
• Select the menu item
Event log
.
• Press ENTER.
A list is displayed containing all error messages since the last time the transmitter was switched on.
• Press
and
to select an error message.
• Press ENTER.
The display indicates the cause of the error.
Special functions\System settings\Event log
Notice!
After reading out the event protocol the error message symbol will be deleted on the display, even if the error has not
been eliminated yet.
The event protocol will be deleted after a restart of the transmitter.