9 Measurement
FLUXUS H721
9.3 Measurement settings
115
UMFLUXUS_H721V1-5EN, 2022-05-15
Measuring range
The sign of the measured value and the measuring range are determined.
• Select
Sign
if the sign of the measured values is to be considered for the output.
• Select
Absolute value
if the sign of the measured values is not to be considered for the output.
• Enter the lowest expected measured value. The unit of measurement of the source item will be displayed.
Start of meas. range
is the value assigned to the value
Output MIN
of the output range.
• Enter the highest expected measured value. The unit of measurement of the source item will be displayed.
End of meas. range
is the value assigned to the value
Output MAX
of the output range.
Last value
20.0 mA
Other value
error value = 3.5 mA
Options\Channel A\Outputs\...\Measured values\Absolute value
Options\Channel A\Outputs\...\Start of meas. range
Options\Channel A\Outputs\...\End of meas. range
Tab. 9.5:
Examples for the error output (output range: 4...20 mA)