9 Measurement
9.3 Measurement settings
FLUXUS H721
2022-05-15, UMFLUXUS_H721V1-5EN
118
Output function test
The function of the output can now be tested.
• Connect an external measuring instrument to the terminals of the output.
• Select
Yes
to test the output. Select
No
to display the next menu item.
• Press ENTER.
• Enter a test value. It has to be within the output range.
• Press ENTER.
If the external measuring instrument displays the entered value, the output functions correctly.
• Select
Repeat
to repeat the test,
Finish
to display the next menu item.
• Press ENTER.
• Select
Yes
to test the status of the output signal. Select
No
to display the next menu item.
• Press ENTER.
• Select a list item as test value.
• Press ENTER.
• Select
Repeat
to repeat the test. Select
Finish
to display the next menu item.
• Press ENTER.
• Press and hold
to return to the main menu.
Options\Channel A\Outputs\...\Test signal
Options\Channel A\Outputs\...\Enter test value
Options\Channel A\Outputs\...\Test mea. range
Options\Channel A\Outputs\...\Enter test value
status value
event value
Status OK
or
Status error
If the external measuring instrument displays the value (min.
output value for
Status error
, max. output for
Status OK
),
the output functions correctly.
Active
or
Passive
If the external measuring instrument displays the value (min.
output value for
Passive
, max. output for
Active
), the output
functions correctly.