![Flexim FLUXUS H721 Operating Instruction Download Page 225](http://html1.mh-extra.com/html/flexim/fluxus-h721/fluxus-h721_operating-instruction_2304735225.webp)
Annex
FLUXUS H721
C Reference
225
UMFLUXUS_H721V1-5EN, 2022-05-15
C.3
Typical properties of selected fluids at 20 °C
C.3.1
Mixtures with non-changing composition
fluid
explanation
sound
speed
[m/s]
density
[g/cm
3
]
kinematic
viscosity
[mm
2
/s]
range of
application
[°C]
WMM
(1)
Water
liquid water
1482
0.999
1
0…350
x
Propane
liquefied under pressure
755
500
0.2
-180…+97
Butane
liquefied under pressure,
coolant R-600
929
577
0.3
-135…+152
x
Ammonia
liquefied under pressure,
coolant R-717
1373
610
0.2
-78…+132
x
Methanol
1119
792
0.7
-95…+240
Ethanol
1158
789
1.5
-110…+241
Acetone
1187
791
0.4
-90…+235
R134a FKW
coolant FC
521
1240
0.2
-100…+100
x
R407C FKW
coolant FC
494
1158
0.1
-20…+81
x
R410A FKW
coolant FC
457
1085
0.1
-130…+71
x
R22 FCKW
coolant FC
557
1213
0.1
-150…+90
x
BP Transcal LT
thermal oil
1365
876
20
-20…+260
x
BP Transcal N
thermal oil
1365
876
94
0…320
x
Shell Thermia B
thermal oil
1365
863
89
0…310
x
Mobiltherm 594
thermal oil
1365
873
7.5
-44…+260
x
Mobiltherm 603
thermal oil
1365
859
55
0…320
x
Gasoline
hydrocarbon with 58 °API
1252
741
1.1
-50…+450
Diesel
hydrocarbon with 38 °API
1380
831
2.2
-50…+450
Jet A1
hydrocarbon with 44 °API
1358
821
1.7
-50…+450
(1)
thermal energy rate coefficient included in the fluid data set