19 Advanced functions
19.5 Modification of the limit for the inner pipe diameter
FLUXUS H721
2022-05-15, UMFLUXUS_H721V1-5EN
182
19.4.3
Information concerning snaps
• Select the menu item
Snap info
.
• Press ENTER.
The following information is displayed:
19.4.4
Deletion of snaps
• Select the menu item
Delete snaps
.
• Press ENTER.
• Select
Yes
or
No
.
• Press ENTER.
19.5
Modification of the limit for the inner pipe diameter
It is possible to modify the lower limit of the inner pipe diameter for a given transducer type.
• Select the program branch
Special functions
.
• Press ENTER.
• Select
Pipe diameter MIN
.
• Press ENTER.
It is possible to define a min. pipe diameter for all relevant transducer frequencies.
• Select
Default
if no customized inputs are to be made and the default values are to be used.
• Press ENTER.
• Select
Customized
if a min. pipe diameter is to be defined.
• Press ENTER.
• Enter the pipe diameter in mm.
• Press ENTER.
Special functions\Snap\Snap info
display
description
Stored snaps:
number of stored snaps
Snaps left:
number of snaps that can still be stored
Ringbuffer:
snap ringbuffer activated
Special functions\Snap\Delete snaps
Special functions\Pipe diameter MIN
Notice!
If a transducer is used below its recommended inner pipe diameter, a measurement might be impossible.