9 Measurement
9.3 Measurement settings
FLUXUS H721
2022-05-15, UMFLUXUS_H721V1-5EN
116
Terminal assignment
The terminals for the connection of the output are displayed.
By pressing
or
further information is displayed.
• Press ENTER.
Output function test
The function of the output can now be tested.
• Connect an external measuring instrument to the terminals of the output.
• Select
Yes
to test the output. Select
No
to display the next menu item.
• Press ENTER.
• Enter a test value. It has to be within the output range.
• Press ENTER.
If the external measuring instrument displays the entered value, the output functions correctly.
• Select
Repeat
to repeat the test,
Finish
to display the next menu item.
• Press ENTER.
• Select
Yes
to test the assignment of the measured value to the output signal. Select
No
to display the next menu item.
• Press ENTER.
• Select in the scroll list
Active
or
Passive
.
• Press ENTER.
If the external measuring instrument displays the value (min. output value for
Passive
, max. output for
Active
), the
output functions correctly.
• Select
Repeat
to repeat the test. Select
Finish
to display the next menu item.
• Press ENTER.
Options\Channel A\Outputs\...\Output info
Options\Channel A\Outputs\...\Test signal
Options\Channel A\Outputs\...\Enter test value
Options\Channel A\Outputs\...\Test mea. range
Options\Channel A\Outputs\...\Enter test value