![Flexim FLUXUS H721 Operating Instruction Download Page 40](http://html1.mh-extra.com/html/flexim/fluxus-h721/fluxus-h721_operating-instruction_2304735040.webp)
6 Mounting
6.2 Transducers
FLUXUS H721
2022-05-15, UMFLUXUS_H721V1-5EN
40
• Place the tension strap around the pipe and insert it into the clasp (see Fig. 6.19).
• Tighten the tension strap.
• Tighten the clasp screw.
Ratchet clasp
• Insert the tension strap into its clamp and the metal spring (see Fig. 6.21). The metal spring does not have to be mounted on:
– steel pipes
– pipes with an outer diameter < 80 mm
– pipes that are not subjected to significant temperature fluctuations
• Position the ratchet clasp, tension strap clamp and metal spring (if necessary) on the pipe (see Fig. 6.17):
– On a horizontal pipe, mount the tension strap clamp laterally to the pipe, if possible.
– Mount the metal spring (if necessary) opposite the tension strap clamp.
• Place the tension strap around the pipe and pass it through the slot of the clasp screw (see Fig. 6.22).
• Tighten the tension strap.
• Cut off the protruding tension strap (see Fig. 6.22).
• Tighten the clasp screw.
Fig. 6.18:
Tension strap with clamp
Fig. 6.19:
Tension strap with quick release
clasp and metal spring
1 – tension strap clamp
1 – clasp screw
2 – metal spring
Fig. 6.20:
Tension strap with band clamp
clasp
1 – clasp screw
Caution!
The edge of the tension strap is very sharp.
Risk of injury.
→
Debur sharp edges.
→
Wear the required personal protective equipment.
→
Observe the applicable rules.
Notice!
In order to release the screw and the tension strap, press the lever down (see Fig. 6.22).
1
2
1
1