![Flexim FLUXUS H721 Operating Instruction Download Page 19](http://html1.mh-extra.com/html/flexim/fluxus-h721/fluxus-h721_operating-instruction_2304735019.webp)
3 General principles
FLUXUS H721
3.4 Undisturbed flow profile
19
UMFLUXUS_H721V1-5EN, 2022-05-15
disturbance: double elbow out of plane (with elbow distance l
2
≥
3 D)
inlet: l
≥
10 D
(l
≥
2 D with disturbance correction)
outlet: l
≥
3 D
disturbance: double elbow out of plane (directly coupled)
inlet: l
≥
40 D
(l
≥
2 D with disturbance correction)
outlet: l
≥
3 D
disturbance: valve
inlet: l
≥
40 D
outlet: l
≥
5 D
disturbance: reducer
inlet: l
≥
10 D
outlet: l
≥
3 D
disturbance: pump
inlet: l
≥
20 D
Tab. 3.3:
Recommended distance from disturbances
D – nominal pipe diameter at the measuring point
l – recommended distance between disturbance and transducer position
l
l
l
2
l
2
l
l
l
l
l
l
l