![Flexim FLUXUS H721 Operating Instruction Download Page 109](http://html1.mh-extra.com/html/flexim/fluxus-h721/fluxus-h721_operating-instruction_2304735109.webp)
9 Measurement
FLUXUS H721
9.3 Measurement settings
109
UMFLUXUS_H721V1-5EN, 2022-05-15
9.3
Measurement settings
• Select the program branch
Options
.
• Press ENTER.
• Select the channel for which a physical quantity has to be
entered (here:
Channel A)
• Press ENTER.
This display will not be indicated if the transmitter has only
one measuring channel.
• Select the menu item
Measurement
.
• Press ENTER.
9.3.1
Selection of the physical quantity
The available physical quantities are displayed in a list.
• Select the physical quantity.
• Press ENTER.
9.3.2
Selection of the unit of measurement
For the selected physical quantity (except sound speed), a scroll list with the available units of measurement is displayed.
• Select the unit of measurement of the physical quantity.
• Press ENTER.
9.3.3
Input of the damping factor
Each displayed measured value is a floating average of the last x seconds, with x being the damping factor. If 0 s is
entered as damping factor, no average is calculated.
The value of 10 s is appropriate for normal flow conditions. If the values fluctuate strongly, caused by a higher dynamic
flow, a higher damping factor can be very useful.
• Enter the damping factor.
• Press ENTER.
Options
Channel x
Measurement
Units of measurement
Physical quantity
Flow velocity
Unit of measurement
Volumetric flow rate
Damping
Mass flow rate
Dynamic damping
Error delay
see annex A
Options\Channel A
Options\Channel A\Measurement\Physical quantity
Options\...\Volumetric flow rate
Notice!
If the physical quantity or the unit of measurement is changed, the settings of the outputs have to be checked.
Options\...\Damping