20 Settings
FLUXUS H721
20.2 Measuring modes
193
UMFLUXUS_H721V1-5EN, 2022-05-15
Toggle time
The toggle time indicates the time interval after which the transmitter toggles between the different channels during the
measurement. The default value is 3 s.
• Press ENTER until the menu item
Toggle time
is displayed.
• Enter a different toggle time value if the default value is no to be used.
• Press ENTER.
This value remains stored until a new toggle time is entered.
Primary display value
• Press ENTER until the menu item
Primary display value
is displayed.
• Select
Flow quantity
to display the selected physical quantity value as primary value during the measurement. Select
Totalizer
to display the totalizer value as primary value during the measurement.
• Press ENTER.
Switching off the display backlight
• Press ENTER until the menu item
Light autom. off
is displayed.
• Select
Yes
to activate the automatic switch-off.
• Press ENTER.
If the automatic switch-off of the display backlight is activated, the backlight is switched off after about 30 s. When
pressing a key or connecting a USB cable, the backlight is switched on again.
20.2
Measuring modes
• Select the menu item
Measurement
in the program branch
Special functions
.
• Press ENTER.
• Select the menu item
Measurement modes
.
• Press ENTER.
Enabling the HPI measurement
• Select
On
to enable the HPI measurement and
Off
to disable it.
• Press ENTER.
• Enter the temperature according to the local reference conditions (default value: 20 °C).
• Press ENTER.
• Enter the pressure according to the local reference conditions (default value: 1 bar).
• Press ENTER.
For further settings, see section 9.1.
Special functions\Dialogs/Menus\Toggle time
Special functions\Dialogs/Menus\Primary display value
Special functions\Dialogs/Menus\Light autom. off
Special functions\Measurement\Measurement modes
Special functions\Measurement\Measurement modes\HPI measurement
Special functions\Measurement\Measurement modes\Reference temp.
Special functions\Measurement\Measurement modes\Reference pressure