14 Outputs (TF7-*721**-******-******a-*****-**; a = U, V, D)
14.3 Activation of a frequency output as pulse output
FLUXUS H721
2022-05-15, UMFLUXUS_H721V1-5EN
152
• Enter the pulse value.
The unit of measurement will be displayed according to the actual physical quantity.
When the counted physical quantity reaches the entered pulse value, a pulse will be transmitted.
• Press ENTER.
• Enter the pulse width.
The range of possible pulse widths depends on the specification of the instrument (e.g., counter, PLC) that is to be
connected to the output.
• Press ENTER.
Terminal assignment
The terminals for the connection of the output are displayed.
By pressing
or
further information is displayed.
• Press ENTER.
14.3
Activation of a frequency output as pulse output
• Select the list item
Frequency
F1
in the menu item
Outputs
.
• Press ENTER.
• Select the list item
Pulse
in the menu item
Source item.
• Press ENTER.
• Select a list item (here:
Pulse +V
).
• Press ENTER.
• Select a list item:
– 0...1 kHz
– 0...5 kHz
– Other range
• Press ENTER.
If
Other range
is selected, enter a value for
Output MAX
.
• Enter the number of pulses per unit.
• Press ENTER.
The unit of measurement will be displayed according to the actual physical quantity.
Options\Channel A\Outputs\Binary B1\Source item\Pulse\Pulse value
Options\Channel A\Outputs\Binary B1\Source item\Pulse\Pulse width
Options\Channel A\Outputs\...\Output info
Options\Channel A\Outputs\Frequency F1
Options\Channel A\Outputs\Frequency F1\Source item\Pulse
Options\Channel A\Outputs\Frequency F1\Source item\Pulse\Pulse +V
Options\...\Frequency F1\Source item\Pulse\F1 Output range
Options\...\Frequency F1\Source item\Pulse\F1 Output range\Pulses per unit