10 Troubleshooting
10.6 Problems with the totalizers
FLUXUS H721
2022-05-15, UMFLUXUS_H721V1-5EN
134
10.6
Problems with the totalizers
The values of the totalizers are too small.
• One of the totalizers has reached the upper limit and has to be reset to zero manually.
The sum of the totalizers is not correct.
• The sum of both totalizers (throughput
Q
) transmitted via an output is not valid after one of the totalizers has overflowed
for the first time.
An interrogation point is displayed after the value of the totalizer.
• The measurement was temporarily impossible, therefore the totalizer value can be wrong.