Summary of Contents for 15409709

Page 1: ...Accu500 Ultrapure and RO water system USER MANUAL...

Page 2: ...able to the Following Accu500 Models 15409709 ACCU500 Water System 10L H with dispenser set 15419709 ACCU500 Water System 15L H with dispenser set 15429709 ACCU500 Water System 20L H with dispenser se...

Page 3: ...Start Up and Time Setting 31 3 3 System Start Up 35 3 4 Routine Operation 40 4 MAINTENANCE 42 4 1 Replacing the Prefilters 43 4 2 Replacing AccuDuo H Pack and AccuDuo U Pack 44 4 3 Replacing Final Fi...

Page 4: ...the continuous flow of high quality ultrapure water Please contact us or your local distributor if you encounter any issues during installation and use Professional engineers are fully trained to supp...

Page 5: ...Making the unit safer to operate Multi optional pre filtration pack removes most large particles calcium and organic compounds to protect RO membrane to extend their usable life The unique automatic...

Page 6: ...and other certification requirements 1 2 Main Applications Pure water can be used in many areas Here are some typical applications With Ultrapure Water With RO Water Important and critical applicatio...

Page 7: ...in 22 74 cm Water Production Rate Ultrapure water Type I 1 5 L min RO water typically 10 15 20 30 L h at 25 Water Tank Capacity 30 L or 60 L RO Rejection Rate 95 Resistivity of Ultrapure Water 18 2 M...

Page 8: ...lish cartridge to remove the last trace of ions then a 0 2 m final filter to the outlet After powering up the system goes into operation mode when the START button is pressed If RO water produced does...

Page 9: ...ductivity temperature and system operation status System is menu driven and displays status of auto run programs Remote Dispenser A high precision resistivity sensor is used to measure resistivity of...

Page 10: ...r temperature compensation Displays The Main system Dispenser Dot matrix backlit LCD display Color LCD Conductivity or resistivity alarm output Can set output upper limit alarm for conductivity RO and...

Page 11: ...x opener to open packages and cut water tubing A wrench to install prefiltration kit 2 2 Items Included Accu500 main system included the following items 1 One User Manual 2 Quality Certificate 3 Acces...

Page 12: ...erbrand 12 2 3 System Structure and Flow Diagrams 2 3 1 System Front View 2 3 2 System Control Panel Display Panel Control Panel RO Outlet Power Indicator Setting Indicator RO Indicator Start Standby...

Page 13: ...after power is on RO indicator Above RO Button It turns solid GREEN when dispensing RO water RO indicator light is also used to indicate whether RO water produced meets preset quality parameters If R...

Page 14: ...em Back Side View In from Tap In from Tank Out to Tank Spare RO Drain Socket for Tank Level Sensor Socket for Dispenser data cable Out to Dispenser Out to System RO Drain FROMTANK TOTANK DRAIN SPARE T...

Page 15: ...Membrane 2 G Check Valve 2 H Flow Limiting Valve 3 I Conductivity Resistivity Sensor 4 J Water Storage Tank 1 K Manual Valve 1 N UV lamp 3 0 H Pack U Pack 1 P Tank Vent Filter 1 S Pressure regulator 1...

Page 16: ...el Sensor UV Lamp Ultrapure Water 1 4 inch tubing Drain Water 1 4 inch tubing Pure Water 1 4 inch tubing Pure Water 3 8 inch tubing Feed Water 3 8 inch tubing TapWater 3 8 inch tubing INLET OUTLET Tan...

Page 17: ...www eu fishersci com fisherbrand 17 2 3 6 Water Dispenser Dispenser button UP outlet Display panel Control Pad Data port Tubing to dispenser head Out to system In from system...

Page 18: ...culation Press 1 and 2 simultaneously to enter user menu 2 In Ready mode press 2 to start volumetric dispensing setting Press 1 and 2 simultaneously to enter user menu 3 In Ready mode press 3 to relea...

Page 19: ...ports without unlocking the connector first or use excessive force to pull the stoppers out as this will damage the connector rendering it useless Please follow the instruction below to remove the sto...

Page 20: ...Remove the remote dispenser from package and assemble it on a stable bench c Take out the cable tubing bundle from the package d Connect the 1 4 tubing in the bundle to the dispenser and the other en...

Page 21: ...rk 1 Install the 3 way ball valve to the tap water inlet Note Fitting is based on 1 2 inch diameter tubing If the tap water outlet is not 1 2 inch you need an adaptor to convert it to 1 2 inch 2 Insta...

Page 22: ...urities which might exist in the pack 6 Block the water outlet with your finger during the process to check whether the assembly leaks 7 Turn off the tap water Connect the outlet PE tubing to the inle...

Page 23: ...et of Prefiltration System and pure water system by 3 8 inch tubing with proper length E Remove the caps from the cartridges and ports of system Wet the O rings on the cartridge with pure water F Gent...

Page 24: ...dge with the lock key to the notch on the look pin I Pressure Relief Open venting ball valve connected with air outlet then turns the tap water on gently to flush out air in PreFilter Cartridge Close...

Page 25: ...uence as indicated in the picture below Never reverse the positions of these packs 1 Open the right side panel of the system 2 Remove blue caps on the new cartridges 3 Install H pack to the LEFT side...

Page 26: ...lock pin till tight 6 Lock the cartridge with the lock key to the notch on the look pin 7 Install the U Pack cartridge to the RIGHT side of the H Pack in the same Insert the lock key into the notch on...

Page 27: ...ew Tank Base Backside View 1 Remove end caps on all ports 2 Take the 1 4 inch 6 35 mm tubing and 3 8 inch 9 53 out from the system accessories bag cut a piece in proper length of each tubing for the t...

Page 28: ...ilter to the top of the PE tank see picture Tank Top View 6 Plug the wire from the liquid level sensor on the tank to the Sensor inlet at the back of the system Tighten the cap 7 Install the faucet a...

Page 29: ...rand 29 Note DO NOT install the final capsule filter The attached final filter will be installed after system initial start up and degassing in the next chapter End of System Installation Proceed to S...

Page 30: ...TDS 1000 ppm Water temperature 5 35 Pressure 1 0 6 0 bar 15 90 psi Water hardness 60 ppm as CaCO3 If water hardness is greater than 100ppm Fisher Scientific suggests you to use an external water soft...

Page 31: ...ee rows of information First row Time XX XX XX XX XX In the following order Year Month Day then Hours Minutes Second row RO status XX X S cm 25 RO Water conductivity compensated to 25 Third row System...

Page 32: ...storage tank is full TANK EMPTY Water level inside the PE tank is below the minimum set point Water cannot be dispensed through the system RO DISPENSE RO water dispense mode time displayed is the pre...

Page 33: ...ain instead of the water tank The RO indicator light blinks It is normal that RO water is above the upper conductivity limit at system start up It usually lasts a few seconds before the conductivity d...

Page 34: ...than one messages appears System Status Message Back Light Meaning EXCH U PACK Yellow H Pack and U Pack cartridges have reached its expected useful life Exchange the cartridges EXCH FINAL FILTER Yello...

Page 35: ...System When the START button is pressed system enters operational mode after an automatic flush of the RO membrane for 180 seconds System automatically detects the quality of the RO water discharges i...

Page 36: ...e minimum set point Water dispensing is stopped automatically System will display Tank Empty on the display panel Once the water tank is refilled sufficiently water can be dispensed again 3 3 3 Flushi...

Page 37: ...inal filter 5 Registering Final Filters In ready mode press and simultaneously to enter user menu Press to move cursor to FILTER and press to reset final filter life Press and simultaneously again to...

Page 38: ...olume collected in liter 5 Press to return to user menu 6 Press and simultaneously to exit user menu 3 3 7 Setting up Clock on the Main System System time is preset If you need to change the clock of...

Page 39: ...twice to exit the editing mode 3 3 8 Setting up Clock on the Dispenser Press and simultaneously to enter user menu Press to move cursor to TIME Press to move cursor to corresponding number and press...

Page 40: ...dispense RO water press the RO button again to stop dispensing b Dispensing RO water from the water tank 3 4 2 Dispensing ultrapure water from the Dispenser In Ready mode press to start recirculation...

Page 41: ...g mode Press and to adjust the desired amount of water Press to to dispense water After dispensing system will continue to recirculate water for 60 seconds Note If the Dispense button is pressed at th...

Page 42: ...Pack U Pack Protects the system 12 months Reduce in resistivity of ultrapure water Polypropylene PP Prefilter Remove particles 2 4 weeks Reduced feed water flow Softening Cartridge Remove Ca and Mg i...

Page 43: ...ted filters 5 Install new filters the 10 m filter in the upstream housing the activated carbon AC cartridge in the middle housing the 1 m filter in the downstream housing 6 Mount the filter housing ba...

Page 44: ...side system panel d Install the AccuDuo H Pack and U Pack as the procedures described in 2 4 3 e Power on the system press the Start button to initiate system Flush and register the AccuDuo U Pack car...

Page 45: ...amp cover to prevent UV light leakage This cover must be on at ALL TIMEs when a UV lamp is installed Caution Keep the UV lamp straight in and out of the stainless steel chamber during its installation...

Page 46: ...rtz glass of the UV lamp 7 Carefully insert the new UV lamp into its chamber When about 2 3 of the lamp is inserted hold the UV lamp and connect it to the ballast cable connector 4 pin connector as sh...

Page 47: ...nt of water If you still feel uncomfortable please seek medical attention Caution RO reverse osmosis membrane should be replaced by a fully trained professional technician 1 Press the START button to...

Page 48: ...th water tubing and then pull tubing out in a gentle manner 6 Connect labeled tubing inside the system to the matching labels RO IN RO OUT and RO DRAIN on the new membrane housing 7 Install the assemb...

Page 49: ...Start button to switch system to standby mode turn off the system and unplug the power cord from the system 2 Put the head of a small head screw driver in the notch 3 Use the back wall of the power c...

Page 50: ...bing connected to the From Tank port put it into the sink Note Please check local regulations on how to properly discharge 0 1 M NaOH 6 Open the blue valve underneath the tank drain all solution 7 Clo...

Page 51: ...ace AccuDuo H and U Pack cartridges Low Pressure Alarm No inlet water Check tap water faucet Prefilter clogged Replace prefilters in the Prefiltration pack Pressure at the tap water too low Call servi...

Page 52: ...d Replace pre filters Reverse osmosis RO membrane clogged Replace RO membrane Final filter clogged Replace final filter Water tank empty Wait till the water tank is filled UP Resistivity dropped at di...

Page 53: ...Accu500 Systems Catalog Number Accu500 System 15409709 ACCU500 Water System 10L H with dispenser set 15419709 ACCU500 Water System 15L H with dispenser set 15429709 ACCU500 Water System 20L H with dis...

Page 54: ...carbon cartridge 10 in 12 pk Pack 15499639 Softener cartridge Cation softener cartridge 10 inches 12 pk Pack 15469669 Reverse osmosis membrane with housing RO membrane for Accu500 10 20 systems 1 pk...

Page 55: ...V lamp 230V Set RAPRC0102B Tank Sanitization Module Tank Sanitization Module with UV lamp 110V Set 15439639 Tank UV Lamp Tank UV Lamp 254nm Each 15479679 30L PE Pure Water Tank 30L PE Pure Water Tank...

Page 56: ...oster Pump for Accu500 10 RO Booster Pump for Accu500 15 RO Booster Pump for Accu500 20 RO Booster Pump for Accu500 30 UP circulation pump Resistivity Sensor Conductivity Sensor Low Pressure switch Pr...

Page 57: ...erials and workmanship when used in compliance with instructions and operating conditions specified in this manual Thermo Fisher Scientific warrants the system for 24 months from the earlier of 1 The...