9.8 Manual Zero
If zero weight cannot be obtained when unloaded, press
[Zero]
.
Zero Indicator
lights up when instrument is at
zero weight status.
9.9 Tare Functions
9.9.1 Manual and automatic tare
This instrument is equipped with the below various tare functions. These functions can be selected in F12.
a.
Off: - Manual Tare
b.
Auto: - Instrument will assume the first stable weight (≥ 20d or 20d
1
) applied is a container and will then
tare off the weight of it. When container is removed and gross weight result is zero, tare effe ct will be
cancelled.
c.
Contin (continuous auto tare): - All stable weight (≥ 20d or 20d
1
) applied will be tared off. When all loads
are removed and gross weight result is zero, tare effect will be cancelled.
Notes: -
•
If Contin is set, select also d.t. (delay time, 0.0 ~ 9.9 second
.
Default = 1.5 second
).
•
Delay time is the time duration from when a stable weight is detected until it is tared off.
•
Enter the preferred d.t. value through numeric keys and press
[Print/M+]
to save.
9.9.2 Repetitive / multiple tare
This instrument is equipped with multiple tare operation capability. Settings: -
•
Set F13 = On to allow multiple tare operation. If it is the case, instrument will permit multiple tare
operation and tare effect can only be cancelled when there is no load on the Platter.
•
Set F13 = Off to disable multiple tare operation. If this is the case, tare effect can only be cancelled
when container is removed and gross weight is zero.
9.9.3 Preset tare
Preset tare allows a pre-determined tare weight value can be entered via numeric keys.
9.9.3.1 Introduce a preset tare value
Enter the pre-determined tare weight value through numeric keys and press
[Print/M+]
to enter. After the pre-
determined tare value has been entered: -
•
Instrument displays the preset tare value entered.
•
Net indicator
appears to indicate the value being displayed is net weight.
9.9.3.2 Cancel a preset tare value
A preset tare value can be cancelled by either one of the below methods: -
•
Remove all loads from platter and press
[Tare]
.
•
Entering a zero preset tare value and press
[Print/M+]
.
9.10 Auto Power Saving & Auto Power off Time
This instrument is equipped auto power saving and auto power off time function. Refer to
F7
on paragraph
6.5
for details.