Faithtech FT68200 Operation Manual
50
3.17. Sequence Test
The sequence test function is used to simulate high-speed and accurate current
waveforms. The FT68200 series load provides 20 sequence files, and each file supports up
to 100 steps. In order to simplify the editing process and improve the operating speed and
accuracy, the sequence file runs in the constant current high range, that is, CCH mode. The
sequence function uses the highest slope when changing the load current. When editing a
sequence single step, the user only needs to set the load current and single step running
time. The sequence single step time range is 0.0001
~
99999s.
Sequence support cycle times and linked files:
Cycle: The number of cycles that the sequence file runs, ranging from 0 to 60000
times. When set to 0, it means that the sequence file is an infinite loop.
Link: After the sequence file has finished running, continue to run the next
sequence file. The link file is set to 0, which means there is no link, that is, the
sequence operation will be stopped after the current file has finished running.
Edit Sequence File
1. Press
(Menu) to go to main menu;
2. Select “Edit”->“Sequence”, press
to go to sequence file select interface;
3. Rotate the knob or input a number to select a sequence file for editing, press
to go to sequence editing;
Figure 3-17 Sequence file edit
Summary of Contents for FT68200 Series
Page 9: ...Faithtech FT68200 Operation Manual 3 1 3 2 7U 8kW 10kW 12kW...
Page 10: ...Faithtech FT68200 Operation Manual 4 1 3 3 11U 14kW 16kW 18kW...
Page 11: ...Faithtech FT68200 Operation Manual 5 1 3 4 14U 20kW 22kW 24kW...
Page 12: ...Faithtech FT68200 Operation Manual 6 1 3 5 20U 30kW 36kW...
Page 13: ...Faithtech FT68200 Operation Manual 7 1 3 6 26U 42kW 48kW...
Page 14: ...Faithtech FT68200 Operation Manual 8 1 3 7 32U 54kW 60kW...