Faithtech FT68200 Operation Manual
75
4. Rotate the knob or input number to set the length, press
to confirm,
cursor moves to “Auto Step”;
5. Input the step number, press
to confirm, cursor moves to “mode”;
6. Select a mode from CC, CV, CR, CP, press
to confirm, cursor moves to
“Step-Time”;
7. “Step-Time” is used to control the running time of the test step, the range is
0.1
~
25.5s. If it is set to 25.5, the actual operation of this step is not restricted by
the "Step-Time", but is determined by the trigger signal: After receiving the
trigger signal, it switch to the next step. After set the “Step-Time”, press
to confirm, cursor moves to “I-Set” (Note: If in procedure 6, CV mode is selected,
then cursor moves to next step “V-Set”; if CR mode is selected, then cursor
moves to next step “R-Set”, if CP mode is selected, then cursor moves to next
step “P-Set”);
8. Set the current value (or voltage, or resistance, or power), press
to
confirm, cursor moves to next step “Short”;
9. "Short" is used to control whether the present step simulates short-circuit
operation. If it is selected as on, it will simulate short-circuit operation in the set
load mode. If it is selected to be off, the load will be carried out according to the
set load value. Press
to confirm, and the focus will automatically move to
the next item "Check";
10. "Check" is used to specify which input parameter is checked by the test step, and
one of Off/Voltage/Current/Power can be selected. Selecting Off means that the
test step does not perform specification check, and the result of this step is Pass.
Press the
key to confirm, and the cursor will automatically move to the
next "Max_trip";
Summary of Contents for FT68200 Series
Page 9: ...Faithtech FT68200 Operation Manual 3 1 3 2 7U 8kW 10kW 12kW...
Page 10: ...Faithtech FT68200 Operation Manual 4 1 3 3 11U 14kW 16kW 18kW...
Page 11: ...Faithtech FT68200 Operation Manual 5 1 3 4 14U 20kW 22kW 24kW...
Page 12: ...Faithtech FT68200 Operation Manual 6 1 3 5 20U 30kW 36kW...
Page 13: ...Faithtech FT68200 Operation Manual 7 1 3 6 26U 42kW 48kW...
Page 14: ...Faithtech FT68200 Operation Manual 8 1 3 7 32U 54kW 60kW...