![Faithtech FT68200 Series User Manual Download Page 62](http://html1.mh-extra.com/html/faithtech/ft68200-series/ft68200-series_user-manual_3224334062.webp)
Faithtech FT68200 Operation Manual
56
Parameter Explanation
Name
Explanation
Von Volt
OPP test turn-on voltage
Von Delay
Delay load time after the input voltage reaches the
turn-on voltage, range: 0.0
~
60.0 seconds
Orig Power
OPP test initial load power
Rise Power
Increment power value for each step
Delay
Running time per step, range: 0.01
~
3600.0 seconds
End Power
Allowed maximum load power
End Volt
Stop test if input voltage is lower than this figure
Max Trip
Judgment upper limit for test power result
Min Trip
Judgment lower limit for test power result
Note
:
If the input voltage is still higher than the end voltage after the end power is
reached, the test stops and the screen displays ERROR.
3.21. Battery Internal Resistance Test
The load uses the DC discharge method to measure the internal resistance of the battery.
According to Ohm's law R=U/I, the load causes the battery to pass two constant currents
in a short time, and the voltage difference between the two ends of the battery is
measured at the same time, and R=(U1-U2)/(I2-I1) is calculated to obtain the internal
resistance of the battery.
FT68200 uses a 4-wire system to measure the internal resistance of the battery. The user
not only needs to connect the input current cable, but also the remote voltage sense cable,
as shown in the figure below.
Figure 3-23 Battery Internal Resistance Test Wiring
The load provides a test result judgment function, which can compare the internal
resistance result with the specified upper and lower limits and report "PASS" or "FAIL".
Summary of Contents for FT68200 Series
Page 9: ...Faithtech FT68200 Operation Manual 3 1 3 2 7U 8kW 10kW 12kW...
Page 10: ...Faithtech FT68200 Operation Manual 4 1 3 3 11U 14kW 16kW 18kW...
Page 11: ...Faithtech FT68200 Operation Manual 5 1 3 4 14U 20kW 22kW 24kW...
Page 12: ...Faithtech FT68200 Operation Manual 6 1 3 5 20U 30kW 36kW...
Page 13: ...Faithtech FT68200 Operation Manual 7 1 3 6 26U 42kW 48kW...
Page 14: ...Faithtech FT68200 Operation Manual 8 1 3 7 32U 54kW 60kW...