40i P - Installation/Operation - DRO installation - (20/32)
3.5.2
Probe connection (connector X5)
Either a 5 V or a 24 V probe may be connected.
Characteristics of probe inputs X5:
5V probe input
Typical value 0,25 mA. ? Vin = 5 V.
High threshold (logic state 1) VIH: From +2.4 V DC on.
Low threshold (logic state 0) VIL: Below +0.9 V DC.
Maximum nominal voltage Vimax = +15 V DC.
24 V probe input
Typical value 0,30 mA. ? Vin = 24 V.
High threshold (logic state 1) VIH: From +12.5 V DC on.
Low threshold (logic state 0) VIL: Below +4 V DC.
Maximum nominal voltage Vimax = +35 V DC.
The unit has two probe inputs (5 V DC and 24 V DC) at connector X5.
Depending on the connection being used, it is possible to choose whether it is activated by the up flank or
by the down flank of the probe (see section
DRO:
).
Probe connection. Connector X5
PIN
SIGNAL
DESCRIPTION
1
CHASSIS
Ground connection
2
+5Vout
+5V output
3
PROBE_5
+5V Probe input
4
PROBE_24
+24V Probe input
5
GNDVOUT
GND output.
6
5Vout
5V output
7
---
---
8
---
---
9
GNDVOUT
GND output.