SONET Tabs
SONET/SDH Application
173
Section RX (SONET)
Alarm Analysis
³
SEF
(Severely Errored Framing): A SEF defect indicates that a
minimum of four consecutive errored framing patterns are received.
³
LOF
(Loss Of Frame): A Loss Of Frame alarm indicates that a Severely
Error Framing (SEF) defect on the incoming SONET signal persists for
at least 3 milliseconds.
³
TIM-S
(Trace Identifier Mismatch - Section): The TIM-S defect indicates
that the received J0 Trace doesn’t match the expected message value.
The TIM-S alarm is only available when
Enable TIM-S
check box from
J0 Trace section has been selected.
J0 Trace
³
Received Message
: Displays the received J0 value. The <crc7>
represents the CRC-7 for a 16-bytes format. The last two bytes of a
64-bytes format, <C
R
> and <L
F
>, represent respectively a carriage
return and a line feed.
³
Enable TIM-S
(Trace Identifier Mismatch - Section): Allows enabling
the Trace Identifier Mismatch for the expected message defined.
Enable TIM-S
has to be enabled to give access to the expected trace
format and message.
³
Expected Message
: Allows entering the expected J0 Trace message. J0
value should be ASCII suitable characters. The default message is
EXFO SONET/SDH
for 16 bytes and
EXFO SONET/SDH Analyzer
Section/RS trace test message
for 64 bytes.
³
Expected Format
: Allows the selection of the expected format: 16 or
64 bytes. The default setting is
16 bytes
.
Summary of Contents for FTB-8100 Series
Page 1: ...User Guide FTB 8100 Series for FTB 200 SONET SDH Application...
Page 18: ......
Page 22: ......
Page 24: ......
Page 94: ......
Page 108: ......
Page 178: ......
Page 224: ......
Page 260: ......
Page 312: ......
Page 352: ......
Page 382: ......
Page 412: ......
Page 490: ......
Page 500: ......