OTN Tabs
SONET/SDH Application
153
ODU TX
Alarm Generation
Type
:
The following alarm types are available. The default setting is
ODU-AIS
.
³
ODU-LOFLOM
(ODU - Loss of Frame Loss Of Multiframe): Generates
error continuously in FAS and MFAS of a multiplexed test case. Only
available for mapped ODU1 in ODU2.
³
ODU-AIS
(ODU - Alarm Indication Signal): Generates an all "1"s pattern
in the entire ODUk signal, excluding the frame alignment overhead (FA
OH), OTUk overhead (OTUk OH) and ODUk FTFL.
³
ODU-OCI
(ODU - Open Connection Indication): Generates a repeating
"01100110" pattern in the entire ODUk signal, excluding the frame
alignment overhead (FA OH) and OTUk overhead (OTUk OH).
³
ODU-LCK
(ODU - Locked): Generates a repeating "01010101" pattern in
the entire ODUk signal, excluding the frame alignment overhead (FA
OH) and OTUk overhead (OTUk OH).
³
ODU-BDI
(ODU - Backward Defect Indication): Generates a “1” in the
BDI (byte 3, bit 5) of the PM overhead field continuously.
³
ODU-FSF
(ODU - Forward Signal Fail): Generates a “00000001” pattern
in the FTFL Byte 0 continuously.
³
ODU-BSF
(ODU - Backward Signal Fail): Generates a “00000001”
pattern in the FTFL Byte 128 continuously.
³
ODU-FSD
(ODU - Forward Signal Degrade): Generates a “00000010”
pattern in the FTFL Byte 0 continuously.
³
ODU-BSD
(ODU - Backward Signal Degrade): Generates a “00000010”
pattern in the FTFL Byte 128 continuously.
On/Off
button: The On/Off button is used to activate/deactivate the
selected alarm. This setting is disabled (Off) by default.
Summary of Contents for FTB-8100 Series
Page 1: ...User Guide FTB 8100 Series for FTB 200 SONET SDH Application...
Page 18: ......
Page 22: ......
Page 24: ......
Page 94: ......
Page 108: ......
Page 178: ......
Page 224: ......
Page 260: ......
Page 312: ......
Page 352: ......
Page 382: ......
Page 412: ......
Page 490: ......
Page 500: ......