OTN Tabs
SONET/SDH Application
163
OPU TX
Note:
The Code field is automatically updated when the Injected payload Type is
changed and vice versa.
³
Code
allows entering the code of the payload type. Choices are
00
to
FF
.
³
Binary
allows either displaying the payload code value in binary
(when enabled) or hexadecimal (when disabled). This setting is
disabled by default.
Alarm Generation
Note:
Alarm generation is only available with multiplexed test case only.
³
OPU-MSIM
(Multiplex Structure Identifier Mismatch): OPU-MSIM is an
OPU alarm that is available for multiplexed test case only. The
OPU-MSIM alarm is generated by corrupting the content of the PSI
(bytes 2 to 5 for ODU1 in ODU2).
Note:
OPU-AIS and OPU-CSF are only available on the OPU client signal
(designated as LO in the standard). In this case, OPU-MSIM is not available.
³
OPU-AIS
(OPU - Alarm Indication Signal): The OPU-AIS alarm is
generated by generating the PRBS 2^11-1 pattern.
³
OPU-CSF
(OPU - Client Signal Fail): The OPU-CSF alarm is generated
by setting the bit 1 of the OPUk PSI[2] byte to “1”.
³
On/Off
button: The On/Off button is used to activate/deactivate the
selected alarm. This setting is disabled (Off) by default.