Port Tabs
SONET/SDH Application
117
Port TX (Electrical Interfaces)
Error Injection
Allows manual or automated error injection.
³
Type
:
The following error types are available with both manual and
automated injection mode.
BPV
(DSn) or
CV
(PDH)
EXZ
(Excessive Zeros) is only available with DS1 and DS3 interfaces.
The default setting is
BPV/CV
.
³
Amount
:
Select the amount of error to be generated.
Choices are
1
through
50
. The default setting is
1
.
³
Send
button:
Press
Send
to manually generate error(s) according to
the
Error Type
and the
Amount of Errors
selected.
³
Rate
:
Press
Rate
to select the injection rate for the selected error. The
rate must be within the minimum and maximum values specified. The
default setting is
1.0E-2
.
³
Continuous
:
Continuous
, when activated, generates the selected
error to its theoretical maximum. This setting is disabled by default.
³
On/Off
button: The
On/Off
button is used to activate/deactivate the
selected automated error at the rate specified or continuously when
continuous is enabled. This setting is disabled (Off) by default.