Proline Promag W 500 PROFINET
Operation
Hauser
165
11.5 Adapting the measuring device to the process
conditions
The following are available for this purpose:
• Basic settings using the
Setup
• Advanced settings using the
Advanced setup
11.6 Performing a totalizer reset
The totalizers are reset in the
Operation
submenu:
• Control Totalizer
• Reset all totalizers
Navigation
"Operation" menu → Totalizer handling
‣
Totalizer handling
Control Totalizer 1 to n
Preset value 1 to n
Reset all totalizers
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection / User
entry
Factory setting
Control Totalizer 1 to n
A process variable is selected
in the
Assign process variable
parameter of the
Totalizer
1 to n
submenu.
Control totalizer value.
• Totalize
• Reset + hold
• hold
• Reset + totalize
• totalize
• Hold
Totalize
Preset value 1 to n
A process variable is selected
in the
Assign process variable
parameter of the
Totalizer
1 to n
submenu.
Specify start value for totalizer.
Dependency
The unit of the selected
process variable is
specified for the totalizer
in the
Unit totalizer
parameter.
Signed floating-point
number
0 l
Reset all totalizers
–
Reset all totalizers to 0 and
start.
• Cancel
• Reset + totalize
Cancel
11.6.1 Function scope of the "Control Totalizer" parameter
Options
Description
Totalize
The totalizer is started or continues running.
Reset + hold
The totaling process is stopped and the totalizer is reset to 0.
hold
The totaling process is stopped and the totalizer is set to its defined start value
from the
Preset value
parameter.
Reset + totalize
The totalizer is reset to 0 and the totaling process is restarted.