Proline Promag W 500 PROFINET
Diagnostics and troubleshooting
Hauser
183
Supported status information
Status
Coding (hex)
BAD - Maintenance alarm
0x24
BAD - Process related
0x28
BAD - Function check
0x3C
UNCERTAIN - Initial value
0x4F
UNCERTAIN - Maintenance demanded
0x68
UNCERTAIN - Process related
0x78
GOOD - OK
0x80
GOOD - Maintenance demanded
0xA8
GOOD - Function check
0xBC
Determining the measured value status and device status via the diagnostic behavior
When the diagnostic behavior is assigned, this also changes the measured value status and
device status for the diagnostic information. The measured value status and device status
depend on the choice of diagnostic behavior and on the group in which the diagnostic
information is located.
The diagnostic information is grouped as follows:
• Diagnostic information pertaining to the sensor: diagnostic number 000 to 199
• Diagnostic information pertaining to the electronics: diagnostic number 200 to 399
• Diagnostic information pertaining to the configuration: diagnostic number 400 to 599
• Diagnostic information pertaining to the process: diagnostic number 800 to 999
Depending on the group in which the diagnostic information is located, the following
measured value status and device status are firmly assigned to the particular diagnostic
behavior:
Diagnostic information pertaining to the sensor: diagnostic number 000 to 199
Diagnostic behavior
(configurable)
Measured value status (fixed assignment)
Device diagnosis
(fixed assignment)
Quality
Quality
Substatus
Coding
(hex)
Category
(NE107)
Alarm
BAD
Maintenance
alarm
0x24
F
(Failure)
Maintenance
alarm
Warning
GOOD
Maintenance
demanded
0xA8
M
(Maintenance)
Maintenance
demanded
Logbook entry only
GOOD
ok
0x80
–
–
Off
Diagnostic information pertaining to the electronics: diagnostic number 200 to 399
Diagnostic number 200 to 301, 303 to 399
Diagnostic behavior
(configurable)
Measured value status (fixed assignment)
Device diagnosis
(fixed assignment)
Quality
Quality
Substatus
Coding
(hex)
Category
(NE107)
Alarm
BAD
Maintenance
alarm
0x24
F
(Failure)
Maintenance
alarm
Warning
Summary of Contents for Proline Promag W 500