Electrical connection
Proline Promag W 500 PROFINET
64
Hauser
7.3.3
Integrating the transmitter into a network
This section only presents the basic options for integrating the device into a network.
For information on the procedure to follow to connect the transmitter correctly → 58.
Integrating via the service interface
The device is integrated via the connection to the service interface (CDI-RJ45).
Note the following when connecting:
• Recommended cable: CAT 5e, CAT 6 or CAT 7, with shielded connector (e.g. brand:
YAMAICHI ; Part No Y-ConProfixPlug63 / Prod. ID: 82-006660)
• Maximum cable thickness: 6 mm
• Length of connector including bend protection: 42 mm
• Bending radius: 5 x cable thickness
N
ic
h
t
u
n
te
r
ar
e
ö
ffn
e
n
Nicht unter
Spannung öffnen
Do not open when
energized
Ne pas ouvrir
sous tension
Power
I/O
≤
42 (1.6)
1
A0033703
1
Service interface (CDI-RJ45)
An adapter for RJ45 and the M12 connector is optionally available:
Order code for "Accessories", option
NB
: "Adapter RJ45 M12 (service interface)"
The adapter connects the service interface (CDI-RJ45) to an M12 connector mounted
in the cable entry. Therefore the connection to the service interface can be established
via an M12 connector without opening the device.