Electrical connection
Proline Promag W 500 PROFINET
56
Hauser
7.2.3
Integrating the transmitter into a network
This section only presents the basic options for integrating the device into a network.
For information on the procedure to follow to connect the transmitter correctly → 50.
Integrating via the service interface
The device is integrated via the connection to the service interface (CDI-RJ45).
Note the following when connecting:
• Recommended cable: CAT5e, CAT6 or CAT7, with shielded connector (e.g. brand:
YAMAICHI ; Part No Y-ConProfixPlug63 / Prod. ID: 82-006660)
• Maximum cable thickness: 6 mm
• Length of connector including bend protection: 42 mm
• Bending radius: 5 x cable thickness
Power
open
press
open
press
ESC
E
+
-
I/O 4
Power
I/O 2
I/O 3
≤
42 (1.6)
1
A0033832
1
Service interface (CDI-RJ45)
An adapter for RJ45 and the M12 connector is optionally available:
Order code for "Accessories", option
NB
: "Adapter RJ45 M12 (service interface)"
The adapter connects the service interface (CDI-RJ45) to an M12 connector mounted
in the cable entry. Therefore the connection to the service interface can be established
via an M12 connector without opening the device.
Summary of Contents for Proline Promag W 500
Page 262: ...www addresses endress com...