Electrical connection
Proline Promag W 400 HART
44
Hauser
3. Connect the transmitter.
Connecting the connecting cable to the sensor connection housing
3.
5
7
4
37
E1
E2
GND
E
42 41
A0032103
17
Sensor: connection module
1. Loosen the securing clamp of the housing cover.
2. Unscrew and lift off the housing cover.
3.
NOTICE
For conduit extensions:
‣
Fit O-ring on cable and push it back sufficiently. When inserting the cable, the O-
ring must be located outside the conduit extension.
Push the cable through the cable entry . To ensure tight sealing, do not remove the
sealing ring from the cable entry.
4. Strip the cable and cable ends. In the case of stranded cables, also fit ferrules
5. Connect the cable in accordance with the terminal assignment → 41.
6. Firmly tighten the cable glands.
7.
L
WARNING
Housing degree of protection may be voided due to insufficient sealing of the
housing.
‣
Screw in the screw without using any lubricant. The threads on the cover are
coated with a dry lubricant.
Reverse the procedure to reassemble the sensor.