11.1 Reading the device locking status
Device active write protection:
Locking status
parameter
Operation → Locking status
Function scope of the "Locking status" parameter
Options
Description
None
The access status displayed in the
Access status display
parameter applies
→ 67. Only appears on local display.
Hardware locked
The DIP switch for hardware locking is activated on the main electronics module.
This locks write access to the parameters (e.g. via local display or operating tool).
CT active - defined
parameters
Only available for Promag W.
The DIP switch for custody transfer mode is activated on the I/O module. This locks
write access to defined parameters (e.g. via local display or operating tool).
For detailed information on custody transfer mode, see the Special
Documentation for the device
CT active - all parameters
Only available for Promag W.
The DIP switch for custody transfer mode is activated on the I/O module. This locks
write access to all the parameters (e.g. via local display or operating tool).
For detailed information on custody transfer mode, see the Special
Documentation for the device
Temporarily locked
Write access to the parameters is temporarily locked on account of internal
processes running in the device (e.g. data upload/download, reset etc.). Once the
internal processing has been completed, the parameters can be changed once
again.
11.2 Adjusting the operating language
Detailed information:
• To configure the operating language → 83
• For information on the operating languages supported by the measuring device
11.3 Configuring the display
Detailed information:
• On the basic settings for the local display → 95
• On the advanced settings for the local display → 104
11.4 Reading measured values
With the
Measured values
submenu, it is possible to read all the measured values.