Commissioning
Proline Promag W 400 HART
90
Hauser
Failure mode
Invert output signal
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection / User
entry
Factory setting
Operating mode
–
Define the output as a pulse,
frequency or switch output.
• Pulse
• Frequency
• Switch
Pulse
Assign pulse output
In the
Operating mode
parameter, the
Pulse
option is
selected.
Select process variable for
pulse output.
• Off
• Volume flow
• Mass flow
• Corrected volume
flow
Off
Value per pulse
In the
Operating mode
parameter, the
Pulse
option is
selected and one of the
following options is selected in
the
Assign pulse output
parameter (→ 90):
• Mass flow
• Volume flow
Enter measured value at which
a pulse is output.
Signed floating-point
number
Depends on country
and nominal
diameter
Pulse width
In the
Operating mode
parameter, the
Pulse
option is
selected and one of the
following options is selected in
the
Assign pulse output
parameter (→ 90):
• Mass flow
• Volume flow
Define time width of the
output pulse.
0.05 to 2
000 ms
100 ms
Failure mode
In the
Operating mode
parameter, the
Pulse
option is
selected and one of the
following options is selected in
the
Assign pulse output
parameter (→ 90):
• Mass flow
• Volume flow
Define output behavior in
alarm condition.
• Actual value
• No pulses
No pulses
Invert output signal
–
Invert the output signal.
• No
• Yes
No
Configuring the frequency output
Navigation
"Setup" menu → Pulse/frequency/switch output 1 to n
‣
Pulse/frequency/switch output 1
to n
Operating mode
Assign frequency output
Minimum frequency value