Proline Promag W 400 HART
Diagnostics and troubleshooting
Hauser
131
Available status signals
Configuration as per HART 7 Specification (Condensed Status), in accordance with
NAMUR NE107.
Symbol
Meaning
A0013956
Failure
A device error is present. The measured value is no longer valid.
A0013959
Function check
The device is in service mode (e.g. during a simulation).
A0013958
Out of specification
The device is being operated:
• Outside its technical specification limits (e.g. outside the process temperature range)
• Outside of the configuration carried out by the user (e.g. maximum flow in parameter
20 mA value
)
A0013957
Maintenance required
Maintenance is required. The measured value is still valid.
A0023076
Has no effect on the condensed status.
12.7 Overview of diagnostic information
The amount of diagnostic information and the number of measured variables affected
increase if the measuring device has one or more application packages.
In the case of some items of diagnostic information, the status signal and the
diagnostic behavior can be changed. Change the diagnostic information → 130
Diagnostic
number
Short text
Remedy instructions
Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
Diagnostic of sensor
043
Sensor short circuit
1. Check sensor cable and sensor
2. Execute Heartbeat Verification
3. Replace sensor cable or sensor
S
Warning
082
Data storage
1. Check module connections
2. Contact service
F
Alarm
083
Memory content
1. Restart device
2. Restore HistoROM S-DAT backup
('Device reset' parameter)
3. Replace HistoROM S-DAT
F
Alarm
170
Coil resistance
Check ambient and process
temperature
F
Alarm
180
Temperature sensor
defective
1. Check sensor connections
2. Replace sensor cable or sensor
3. Turn off temperature
measurement
F
Warning
181
Sensor connection
1. Check sensor cable and sensor
2. Execute Heartbeat Verification
3. Replace sensor cable or sensor
F
Alarm
Diagnostic of electronic
201
Device failure
1. Restart device
2. Contact service
F
Alarm
Summary of Contents for Proline Promag W 400 HART
Page 197: ......