11.6.2 Function scope of the "Reset all totalizers" parameter
Options
Description
Cancel
No action is executed and the user exits the parameter.
Reset + totalize
Resets all totalizers to 0 and restarts the totaling process. This deletes all the flow
values previously totalized.
11.7 Showing data logging
The
Extended HistoROM
application package must be enabled in the device (order
option) for the
Data logging
submenu to appear. This contains all the parameters for the
measured value history.
Data logging is also available via:
• Plant Asset Management Tool FieldCare → 76.
• Web browser
Function range
• A total of 1000 measured values can be stored
• 4 logging channels
• Adjustable logging interval for data logging
• Display of the measured value trend for each logging channel in the form of a chart
A0016222
34
Chart of a measured value trend
• x-axis: depending on the number of channels selected displays 250 to 1000 measured
values of a process variable.
• y-axis: displays the approximate measured value span and constantly adapts this to the
ongoing measurement.
If the length of the logging interval or the assignment of the process variables to the
channels is changed, the content of the data logging is deleted.
Navigation
"Diagnostics" menu → Data logging
‣
Data logging
Assign channel 1...4
Logging interval
Clear logging data
Data logging
Logging delay