Diagnostics and troubleshooting
Proline Promag W 400 HART
132
Hauser
Diagnostic
number
Short text
Remedy instructions
Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
242
Software incompatible
1. Check software
2. Flash or change main electronics
module
F
Alarm
252
Modules incompatible
1. Check electronic modules
2. Change electronic modules
F
Alarm
261
Electronic modules
1. Restart device
2. Check electronic modules
3. Change I/O Modul or main
electronics
F
Alarm
262
Sensor electronic
connection faulty
1. Check or replace connection cable
between sensor electronic module
(ISEM) and main electronics
2. Check or replace ISEM or main
electronics
F
Alarm
270
Main electronic failure
Change main electronic module
F
Alarm
271
Main electronic failure
1. Restart device
2. Change main electronic module
F
Alarm
272
Main electronic failure
1. Restart device
2. Contact service
F
Alarm
273
Main electronic failure
Change electronic
F
Alarm
275
I/O module defective
Change I/O module
F
Alarm
276
I/O module faulty
1. Restart device
2. Change I/O module
F
Alarm
283
Memory content
1. Reset device
2. Contact service
F
Alarm
302
Device verification active Device verification active, please wait. C
Warning
311
Electronic failure
1. Do not reset device
2. Contact service
M
Warning
372
Sensor electronic (ISEM)
faulty
1. Restart device
2. Check if failure recurs
3. Replace sensor electronic module
(ISEM)
F
Alarm
373
Sensor electronic (ISEM)
faulty
1. Transfer data or reset device
2. Contact service
F
Alarm
375
I/O- communication
failed
1. Restart device
2. Check if failure recurs
3. Replace module rack inclusive
electronic modules
F
Alarm
376
Sensor electronic (ISEM)
faulty
1. Replace sensor electronic module
(ISEM)
2. Turn off diagnostic message
F
Warning
377
Sensor electronic (ISEM)
faulty
1. Check sensor cable and sensor
2. Perform Heartbeat Verification
3. Replace sensor cable or sensor
F
Warning
382
Data storage
1. Insert T-DAT
2. Replace T-DAT
F
Alarm
383
Memory content
1. Restart device
2. Delete T-DAT via 'Reset device'
parameter
3. Replace T-DAT
F
Alarm
512
Sensor electronic (ISEM)
faulty
1. Check ECC recovery time
2. Turn off ECC
F
Alarm